Input
Range / Resolution
Accuracy
Sensor type
Sensor life
Loop output
Display
Set point
Setting
Monitoring
Supply Voltage
Cable
Operating Temperature
Sensor Housing
Enclosure Detail
Protection Class
Cabinet Material
Cable Entry
Dimension
Mounting
Weight
|
Oxygen / Toxic gases, Please select from the Table
As specified in the Table
± 1% of the full scale value
Electrochemical sensor, Imported make
Expected 24 months
Industrial standard 4-20mA DC O/P with permissible load 750 ohm at 24 VDC, 25mA
0.5”, 3½ digit LCD Display
Two independent set point, LO & HI with blinking indication with programmable latch & non-latch
Facility only for indication purpose. (There is no relay output)
Setting by 4 keys on front panel i.e. ‘SET / NEXT’, ‘UP’, ‘DOWN’ & ‘TEST / BACK’
I) By Display :Sensor Break/ Open / Over range / Under range Shows “Err” Indication
ii) By loop current : Up scale current : 22mA, Down scale current : 3.6mA to 3.9mA, Selectable
14VDC to 36VDC
1.5 sq. mm 2 Core, Flexible OR Armoured shielded cable
0 to 500C ± 5%, Relative humidity: Below 95% non condensing
P.V.C. / P.P. / TEFLON material / SS304 / SS316, Please specify while ordering
OT - 1000 - WP ( Weatherproof ) OT - 1000 - FLP ( Flameproof )
IP-65 IP-65 for Gas Gr. IIA & IIB, ATEX, CCOE & CMRI Approved
Polycarbonate, gray colour Cast Alluminium alloy, LM-6 / C.I. Enclosure
1 No, Thermoplastic cable gland (IP 68) 1 No. Double compression cable gland (Ex-Proof), ¾” ET type
190mm(H) x 80mm(W) x 55mm(D) 190mm (H) x 185mm (W) x 100mm (D)
Wall Mounting / Stand Mounting Wall Mounting / Stand Mounting
250 Grams 1Kg |